نویسندگان | Mohammad ali Nasseri, |
---|---|
نشریه | Sensors and Actuators B: Chemical |
شماره صفحات | 174-179 |
شماره سریال | 98 |
شماره مجلد | 2 |
ضریب تاثیر (IF) | 5.401 |
نوع مقاله | Full Paper |
تاریخ انتشار | 2004 |
رتبه نشریه | ISI |
نوع نشریه | چاپی |
کشور محل چاپ | هلند |
نمایه نشریه | ISI،JCR،Scopus |
چکیده مقاله
Poly(vinyl chloride) (PVC) membrane electrodes based on two complexes of Schiff base 2,2′[4,4′-diphenylmethane bis (nitromethylidyne)] bisphenol, L, with copper (II) and iron (III) ions were prepared, characterized and used for determination of triiodide ions. The electrodes based on CuL and FeL complexes show an anionic response slope of −60 mV per decade in a linear activity range 6.0×10−1 to 8.0×10−6 mol l−1 and 5.0×10−1 to 1.0×10−5 mol l−1 to triiodide ion with lower detection limits of 4.0×10−6 and 6.0×10−6 mol dm−3, respectively. The proposed electrodes have fast response time (15 s) and their responses are independent of pH of the test solution in the range 3.5–9.0. The electrodes revealed very good selectivities for triiodide ion over a variety of anions. They were used as an indicator electrode in the potentiometric titration of triiodide ions and determination of ascorbic acid in Vitamin C tablet.
tags: : Triiodide ion-selective electrodes; Liquid membrane; Schiff base complexes; Potentiometry