Triiodide PVC Membrane Electrodes Based on Charge-Transfer Complexes

نویسندگان_
نشریهAnalytical Chemistry
شماره صفحات2591-2595
شماره سریال74
شماره مجلد11
نوع مقالهFull Paper
تاریخ انتشار2002
رتبه نشریهISI
نوع نشریهچاپی
کشور محل چاپایران
نمایه نشریهISI،JCR،Scopus

چکیده مقاله

Three new electrodes were prepared by incorporating two different charge-transfer complexes and amino crown ether into plasticized PVC membranes. The electrodes showed Nernstian response to triiodide ion over the activity range from 1.0 × 10-5 to 1.0 × 10-1 mol‚L-1 with detection limits at ∼1.0 × 10- 6 mol‚L-1. The resulting electrodes have fast response times (20-30 s) and good stabilities (4 months) and can be used over a wide pH range of 2.5-9.0. The proposed electrodes exhibit anti-Hofmeister behavior with excellent selectivity toward triiodide ion against a wide range of common interferences. Comparative study suggests that amino (aza) crown ether alone or in the form of a charge-transfer complex with iodine, as an ionophore in a PVC liquid membrane, is sensitive to triiodide ion. The electrodes were used as indicator electrodes in potentiometric titration of triiodide ion against thiosulfate ion.

لینک ثابت مقاله

tags: ion-pair; crown ether; triiodide; potentiometry