| نویسندگان | _ |
| نشریه | Analytical Letters |
| شماره صفحات | 2026-2040 |
| شماره سریال | 42 |
| شماره مجلد | 13 |
| ضریب تاثیر (IF) | 0.965 |
| نوع مقاله | Full Paper |
| تاریخ انتشار | 2009 |
| رتبه نشریه | ISI |
| نوع نشریه | چاپی |
| کشور محل چاپ | ایران |
| نمایه نشریه | ISI،JCR،Scopus |
چکیده مقاله
The potentiometric response characteristics of a new copper(II) ion-selective PVC membrane electrode based on erythromycin ethyl succinate (EES) as ionophore were investigated. The electrode exhibited a Nernstian response to Cu2+ ions over the activity range of 1.5 × 10−2 to 2.0 × 10−6 mol L−1 with a limit of detection of 6.3 × 10−7 mol L−1. Stable potentials were obtained in the pH range of 5.5–6.5. The potentiometric selectivity coefficients were calculated by using fixed interference method and revealed no important interferences except for Ag+. This electrode was successfully applied as an indicator electrode in determination of copper ions in real water samples.
لینک ثابت مقاله