| نویسندگان | Mohammad Mehdi Firoozabadi,N. Ghal-Eh |
| نشریه | Nuclear Instruments and Methods in physics Research A |
| شماره صفحات | 162574-0 |
| ضریب تاثیر (IF) | 1.142 |
| نوع مقاله | Full Paper |
| تاریخ انتشار | 2019 |
| رتبه نشریه | ISI |
| نوع نشریه | چاپی |
| کشور محل چاپ | ایران |
| نمایه نشریه | JCR،Scopus |
چکیده مقاله
The longitudinal response uniformity of the rectangular-shaped plastic scintillators used in a thermal neutron
imaging system proposed by Ghal-Eh and Green in 2016 is vital for its precise measurements in quality
assurance studies of boron neutron capture therapy (BNCT). The MCNPX-PHOTRACK hybrid code has been
used to model the response of rectangular plastic scintillator of partially painted surface in order to reduce as
much as possible the longitudinal non-uniformity which usually exist in large and/or long scintillators. The
results show that the optimum uniformity corresponds to fully-polished surface which coincides with FLUKA
light transport simulations.