Efficiency of tungsten-polymer composite shields on fetal dose reduction in chest CT scans

نویسندگانKeyhandokht Karimi Shahri,Atiyeh Ebrahimi-Khankook,Parisa Akhlaghi,Najmeh Mohammadi,Yasaman Rezaei-Moghaddam
نشریهMEDICAL ENGINEERING & PHYSICS
شماره صفحات104008-104013
شماره سریال118
شماره مجلد1
ضریب تاثیر (IF)1.923
نوع مقالهFull Paper
تاریخ انتشار2023
نوع نشریهالکترونیکی
کشور محل چاپایران
نمایه نشریهISI،JCR،Scopus

چکیده مقاله

Given their desirable shielding properties, 5 polymer composite shields reinforced with tungsten were selected and their effects as gamma shields on fetal dose reduction were investigated. According to the results, the selected shields reduce the fetus brain, the fetus lungs, the fetus kidneys, and the total fetus dose almost 34.17%- 41.19%, 20.47%-25.08%, 9.27%-12.13%, and 15.39%-18.69%, respectively, at tube voltage of 80 kVp. At the higher tube potentials, the values of dose reduction were smaller. Moreover, it was observed that polymers named PHEMA-WO3 and RS-U-30 had an excellent shielding ability among the other studied composites.

لینک ثابت مقاله

tags: Tungsten composites Pregnant phantom Gamma shield Dose reduction Chest CT scan