Validity and reliability of a new non-invasive method for measuring the scoliosis deformity

نویسندگانMohammad Yousefi,Zahra Ebrahimabadi,sanaz Shanbeh zadeh,alye daryabor
نشریهSeries on Biomechanics
شماره صفحات18-26
شماره سریال37
شماره مجلد2
نوع مقالهFull Paper
تاریخ انتشار2023
نوع نشریهچاپی
کشور محل چاپبلغارستان
نمایه نشریهScopus

چکیده مقاله

The objective of the study was to assess the reliability and validity of a new non-invasive system for measuring scoliosis spinal angles in thoracic and lumbar region. Three raters independently measured 27 individuals with scoliotic deformity three times on each subject at a one-week interval. Validity was evaluated by assessing the correlation between the X-ray Cobb angle and new infrared method (Body Landmark Analyzer system (BLA)), and reliability (inter and intra rater) were investigated with raters 1-3. The results of the current study revealed excellent inter- (ICC: 0.81- 0.87) and intra-rater (ICC: 0.87-0.90) reliability for spinal curve angle measurement and high validity with Cobb angle measures of X-ray (ICC: 0.80-0.88) in thoracic region. The findings of this study showed that BLA system could provide a reliable, safe, and valid device to measure scoliotic curve angle

لینک ثابت مقاله

tags: Validity, reliability, scoliosis, deformity, non-invasive method